Consumption electric power of WA500 (A)-Condensation latent
heat (*)(B)=heating value(C)
(A): Heater capacity×860Kcal = 3.8(Kcal)×860
= 3500Kcal/H
(B): Pumped volume of water×water density×latent
heat of water = 5×1×540 = 2700Kcal/H * Underlined
values are quantitative.
(C): 3500 - 2700 = 800Kcal/H
Therefore, the heating value is 800Kcal/H
Heating value is almost zero when the tank is full and the
heater is OFF. No heat is generated at the heat controller.
*Condensation latent heat: Heat is conducted to the
coolant to discharge the heat outside.
|