FAQ on Constant Temperature and Humidity
Chambers (IW221)
|
 |
The average wind velocity inside the chamber is defined as 2.4m/s
in the catalog. Does the range of the wind velocity fall within 0.5
to 2.5m/s? |
 |
No, it does not. According to the wind velocity data measured inside
the bath, the maximum wind velocity reaches 2.9m/s depending on the
position inside the bath. |