FAQ on Tubing Pump System (7531-00,
Others)
|
 |
We want to connect the membrane filter. The max. discharge pressure
of the membrane filter is 2.1kg. Is there any problem if we use 7531-00·L/S16
tube, EASY-LOAD head? We only use 1m of the tube. |
 |
The max. discharge pressure of the tube shown in the catalog is
the pressure limitation that the tube can withstand, and it is assumed
that the notation for the filter is the same. It is predicted
that there is no problem at the beginning of the usage, but once the
membrane filter is clogged, the pressure is loaded to the tube and
the tube might burst in the worst case. Check the flow rate
at the beginning, and replace the filter if the flow rate extremely
decreases. Check the tube periodically and replace it if the tube
is crushed or hardened. Please decide the term of the periodic
check after confirming the level of deterioration similarly
to the membrane filter. |
Inquiry
Home
Copyright(c) 1998 YAMATO SCIENTIFIC CO.,LTD. ALL rights