| This safety cabinet separates the air supply
system and the exhaust system, and adopts the non-circulation
system, "All-Fresh". Furthermore, both inside and outside of the
safety cabinet are isolated with the HEPA filter and the air barrier
at the operation opening to prevent the experimenter from infection,
and to protect the experimental material simultaneously. This
safety cabinet is suitable for the experiments with possibility
of biohazard that should not be performed under the draft circulation
condition. |
| Specifications |
| Model |
Model GB |
| Item |
Safety Cabinet |
| Construction and Material |
Main Body |
Material |
Grind Finish Plate |
| Welded Part |
Grind Finish after Continuous Welding (Work Room) |
| Interior Corner |
R-finish |
| Base |
Steel (Baking Finish of Bark) |
| Packing |
Neoprene Rubber |
| Shut-off Valve |
Enclosed Manual Valve |
| Glass |
Safety Laminated Glass |
| HEPA Filter |
H · 110A×2 Sheets |
| Pressure Display |
Precision Differential Pressure Gauge×1 |
| Fluorescent Lamp |
200W×4 |
| Glove |
Neoprene Rubber |
| Performance |
Dust Collecting Efficiency |
99.99 % or more with particle of 0.3µm |
| No. of Ventilation Times |
40 times/H or more |
| Airtightness |
Leak Test by Freon Gas Pressurizing: 10-7ml/s or more |
| Interior Pressure |
-15mmAq: Displayed with Precision Differential Pressure Gauge |
| Safety Measurement |
Alarm Buzzer ON at Pressure Rising |
|
| Contact
us |
Request reference material Request for documents Return to Home
Copyright(c) 1998 YAMATO SCIENTIFIC CO.,LTD. All rights reserved. For more information, call 0120-405-525(Free dial) E-mail:english-website@yamato- net.co.jp
|