Safety Cabinet
Class III
This safety cabinet separates the air supply system and the exhaust system, and adopts the non-circulation system, "All-Fresh". Furthermore, both inside and outside of the safety cabinet are isolated with the HEPA filter and the air barrier at the operation opening to prevent the experimenter from infection, and to protect the experimental material simultaneously. This safety cabinet is suitable for the experiments with possibility of biohazard that should not be performed under the draft circulation condition.


Specifications
Model Model GB
Item Safety Cabinet
Construction and Material Main Body Material Grind Finish Plate
Welded Part Grind Finish after Continuous Welding (Work Room)
Interior Corner R-finish
Base Steel (Baking Finish of Bark)
Packing Neoprene Rubber
Shut-off Valve Enclosed Manual Valve
Glass Safety Laminated Glass
HEPA Filter H · 110A×2 Sheets
Pressure Display Precision Differential Pressure Gauge×1
Fluorescent Lamp 200W×4
Glove Neoprene Rubber
Performance Dust Collecting Efficiency 99.99 % or more with particle of 0.3µm
No. of Ventilation Times 40 times/H or more
Airtightness Leak Test by Freon Gas Pressurizing: 10-7ml/s or more
Interior Pressure -15mmAq: Displayed with Precision Differential Pressure Gauge
Safety Measurement Alarm Buzzer ON at Pressure Rising