Safety Cabinet
Class IIC
Class IIC
Model SCV-1303ECIIC
This safety cabinet separates the air supply system and the exhaust system, and adopts the non-circulation system, "All-Fresh". Furthermore, both inside and outside of the safety cabinet are isolated with the HEPA filter and the air barrier at the operation opening to prevent the experimenter from infection, and to protect the experimental material simultaneously. This safety cabinet is suitable for the experiments with possibility of biohazard that should not be performed under the draft circulation condition.


Specifications
Product Code 100000 100000 100000  
Model SCV 803ECIIC 1303ECIIC 1903ECIIC  
Dust Collecting Element HEPA Filter
Dust Collecting Efficiency 99.99 % or more with particle of 0.3µm, passed Scan Test
Gas Quantity Exhaust 14.0m3/min. or more 24.5m3/min. or more 36.0m3/min. or more  
Wind Velocity Inside Work Bench (50/60Hz) Average 0.30 to 0.52m/s or more In accordance with the NSF standard
Inflow Opening Average 0.55 to 0.70m/s or more (Standard is 0.5 m/s or more)
Bacteria Test Method Make sampling by spraying the bacillus subtilis spore by nebulizer
Presonal Protection
Test
No. of colonies from AGI sampler synthesis floating liquid: 10 or less
No. of colonies from the slit sampler: 5 or less
Product Protection
Test
No. of colonies on the petri dish placed all over the work bench: 5 or less
Cross · Contamination
Test
No. of colonies on the petri dish placed 355.6 mm or more apart from the side surface: 0 or less
Airtightness of Main Body When the freon gas is applied into the main body, the leakage from parts of the body under pressure of 498 Pa (50.8 mmAq) is 8.9×10-5ml/s or less
Intensity of Ultra Violet Rays Whole Work Bench: 40µW/cm3or more
Power Source 100VAC Single Phase 50Hz/60Hz
Material of Work Bench Equal to Surface finish No.4 of JIS
Corners shall be rounded.
Dimension of Main Body: Width (mm) 1,000 1,500 2,150
D×H (separable) 800×2,300 (1,700)
Contact us


Request reference material

Request for documents

Return to Home

Copyright(c) 1998 YAMATO SCIENTIFIC CO.,LTD. All rights reserved. For more information, call 0120-405-525(Free dial) E-mail:english-website@yamato- net.co.jp