Safety Cabinet
Class IIA

Model SCV-1303ECIIAJ
Japan Air Cleaning Association (Social Foundation) establishes the standard (JACA std. No. 16C) regarding Class II safety cabinet, and executes certification test. This safety cabinet has acquired the authorization.


Specifications
Product Code
Model
100000
SCV-1303ECIIAJ
Dust Collecting Element HEPA Filter
Dust Collecting Efficiency 99.99 % or more with particle of 0.3µm, passed Scan Test
Gas Quantity Exhaust 7.0/8.7m3/min. or more
Wind Velocity Inside Work Bench
(50/60Hz)
Average m/s Back Row     0.295/0.375
Middle Row  0.315/0.415
Front Row     0.465/0.605
Inflow Opening Average m/s 0.45/0.56
Bacteria Test Method Make sampling by spraying the bacillus subtilis spore with nebulizer (on the basis of NSF Standard).
Presonal Protection
Test
No. of colonies from the AGI sampler synthesis floating liquid: 10 or less
No. of colonies from the slit sampler: 5 or less
Product Protection
Test
No. of colonies on the petri dish placed all over the work bench: 5 or less
Cross Contamination
Test
No. of colonies on the petri dish placed 355 mm or more apart from the side surface: 2 or less
Airtightness of Main Body When the freon gas is applied into the main body, the leakage from parts of the body under pressure of 50 mmAq is 5×10-7ml/s or less
Intensity of Ultra Violet Rays Whole Work Bench: 40µW/cm3 or more
Power Source 100VAC Single Phase 50Hz/60Hz (15A×2)
Material of Work Bench Relative to Surface finish No.4 of JIS
Corners shall be rounded.
Dimension of Main Body W1,500×D800×H2,030 (separable) mm
Weight of Main Body Approx. 320kg    