Product Code
Model |
100000
SCV-1303ECIIAJ |
| Dust Collecting Element |
HEPA Filter |
| Dust Collecting Efficiency |
99.99 % or more with particle of 0.3µm, passed Scan Test |
| Gas Quantity |
Exhaust |
7.0/8.7m3/min. or more |
| Wind Velocity |
Inside Work Bench
(50/60Hz) |
Average m/s |
Back Row 0.295/0.375 |
| Middle Row 0.315/0.415 |
| Front Row 0.465/0.605 |
| Inflow Opening |
Average m/s |
0.45/0.56 |
| Bacteria Test |
Method |
Make sampling by spraying the bacillus subtilis spore with nebulizer (on the basis of NSF Standard). |
Presonal Protection
Test |
No. of colonies from the AGI sampler synthesis floating liquid: 10 or less
No. of colonies from the slit sampler: 5 or less |
Product Protection
Test |
No. of colonies on the petri dish placed all over the work bench: 5 or less |
Cross Contamination
Test |
No. of colonies on the petri dish placed 355 mm or more apart from the side surface: 2 or less |
| Airtightness of Main Body |
When the freon gas is applied into the main body, the leakage from parts of the body under pressure of 50 mmAq is 5×10-7ml/s or less |
| Intensity of Ultra Violet Rays |
Whole Work Bench: 40µW/cm3 or more |
| Power Source |
100VAC Single Phase 50Hz/60Hz (15A×2) |
| Material of Work Bench |
Relative to Surface finish No.4 of JIS
Corners shall be rounded. |
| Dimension of Main Body |
W1,500×D800×H2,030 (separable) mm |
| Weight of Main Body |
Approx. 320kg |