Safety Cabinet
Class I
Class I
  Model SCV-1303ECI
This device filters the exhaust gas by HEPA filter to maintain the work space with no-load and prevent leakage of biological material. It passed the bacteria test and the airtightness test on the basis of NSF Standard of the United States.


Specifications
Product Code 100000 100000 100000 100000
Model SCV 803ECI 1303ECI 1603ECI 1903ECI
Dust Collecting Element HEPA Filter
Dust Collecting Efficiency 99.99% or more at 0.3µm, passed Scan Test
Exhaust Gas Quantity (m3/min. or more) 4.9 7.8 9.6 11.7
Inflow Wind Velocity at Supply Opening Average 0.5m/s or more
Bacteria Test Method Make sampling by spraying the bacillus subtilis spore on the basis of NSF Standard.
Presonal  
Protection
Test
No. of colonies from the AGI sampler synthesis floating liquid: 20 or less
No. of colonies from the slit sampler: 5 or less
Airtightness of Main Body When the freon gas is applied into the main body, the leakage from parts of the body under pressure of 498Pa (50.5mmAq) is 8.9×10-5ml/s or less
Intensity of Ultra Violet Rays Whole Work Bench: 40µW/cm3 or more
Power Source 100VAC Single Phase 50Hz/60Hz (15A×2)
Material of Work Bench Relative to Surface finish No.4 of JIS
Dimension of Main Body: Width (mm) 1,000 1,500 1,800 2,150
D×H 800×2,030mm
Weight of Main Body (kg) 260 320 370 400