YI-AT-100/001/002 |
Performance |
| Method |
Forced air circulation type |
| Operating temperature range |
40 to 260deg.C |
| Temperature uniformity |
±3deg.C (at 175deg.C) |
| Temperature rise time |
15 min.(50deg.C --> 175deg.C with products loading) |
| Cooling time |
30 min.(175deg.C --> 50deg.C) |
Configuration |
| Exterior meterial |
Cold-rolled steel sheet with melamine resin baked-painting |
| Interior material |
Stainless steel sheet |
| Heater |
Stainless steel pipe heater with fin, 9kW for the each chamber |
| Motor |
Capacitor motor 200W |
| O2 analizer port |
R1/4 pipe thread (female) |
| N2 gas inlet/CDA inlet port |
I.D.8mm tube connector (with 100旭/min. flow meter)/O.D.8mm tube connector |
Controller |
| Temperature control |
High peaformance program controller UP550 |
| Temperature seisor |
Thermocouple type K (5 point with sensor guard) |
| Temperature recoeder |
One pen hybrid recorder μR1000 (Programmed operation (max. 99 segments, 30 patterns) |
| Signal tower |
Green--Operation, Yellow--Standby, Red--Error with alarm |
| Emergency stop switch |
When set to ON, it stop system operation completely |
| Program details |
Max. 99 segments, 30 patternes |
Standard |
| Internal dimensions(W×D×Hmm) |
550×550×550/chamber |
| Internal capacity |
Approx. 166 liters/chamber |
| Water inlet/outlet port |
O.D. 1/2 inch tube fitting |
| Power source (50/60Hz) |
3 phase AC220V/380V |
Communication |
| External comunucation function |
RS-485 interface |