| Model |
DPB530 |
| System |
Heat radiation and transfer system with vacuum and 4-side hot plates |
| Performance |
Operating temp. range |
Room temperature + 10 to 150deg.C |
| Operating vacuum level range |
101 kPa to 1 Pa or less (at no load), displayed on the vacuum indicator |
| Attainable vacuum level |
1 Pa or less (at no-load), depending on vacuum pump capacity |
| Temp. adjustment accuracy |
±1deg.C (at 100deg.C), both in atmosphere and vacuum, temperatures at five points in chamber |
| Temp. distribution accuracy |
±5deg.C (at 100deg.C), both in atmosphere and vacuum, temperatures at five points in chamber |
| Temp. distribution accuracy of bottom hot plates |
±2deg.C (at 100deg.C), both in atmosphere and vacuum |
| Time to attain max. temp. |
Approx. 90 min. up to 100deg.C, both in atmosphere and vacuum, shortest reach time of temperatures at five points in chamber |
| Open-to-atmosphere time |
One minute or less (at full opened manual valve), adjustable with the manual valve |
| Configuration/Functions |
Exterior |
Chrome free electro galvanized steel plate Chemical proof baking finish |
| Interior |
Stainless steel plate, 2 stages chamber (Vacuum level and temperature adjustable individually) |
| Door |
Material inside chamber: Aluminum, without heat insulator, one touch handle mechanism |
| Door packing |
Vacuum O-ring (Fluororubber) |
| Heater |
3.6 kW/chamber |
| Temperature controller |
PID control with a micro computer (separate control of chambers x 2-system heater control) |
| Temp. control/overheat preventive device sensor |
K-thermocouple |
| Sample temperature sensor |
K-thermocouple 5/chamber |
| Vacuum valve |
Air driven, NW50 |
| Vacuum meter |
101 kPa to 10-2 Pa, vacuum indicator |
| Safety unit |
Over current ELB, independent overheat preventive device, stop valve, emergency relief valve, vacuum valve filter regulator |
| Standards |
Internal size |
W512 x D315 x H155 mm/chamber |
| External size |
W1100 x D1100 x H1400 mm |
| Bottom surface effective size |
W380 x D300 x H155 mm |
| Exhaust port |
NW50 flange |
| Air supply port |
Rc1/2 |
| Power supply |
3 phase AC200V 50/60 Hz 22A (excluding vacuum pump power supply) |
| Compressed air |
0.5 Mpa or more, φ6 mm tube coupler is used. |