| Model |
DPB430 |
| System |
Heat radiation and transfer system with vacuum and 4-side hot plates |
| Performance |
Operating temp. range |
Room temperature + 10 to 150deg.C |
| Operating vacuum level range |
101 kPa to 0.1 kPa |
| Attainable vacuum level |
1 Pa or less (at no-load), depending on vacuum pump capacity |
| Temp. adjustment accuracy |
±2deg.C (at 100deg.C), both in atmosphere and vacuum, temperatures at five points in chamber |
| Temp. distribution accuracy |
±5deg.C (at 100deg.C), both in atmosphere and vacuum, temperatures at five points in chamber |
| Temp. distribution accuracy of bottom hot plates |
±2deg.C (at 100deg.C), both in atmosphere and vacuum |
| Time to attain max. temp. |
Approx. 100 min. up to 100deg.C, both in atmosphere and vacuum, shortest reach time of temperatures at five points in chamber |
| Configuration/Functions |
Exterior |
Chrome free electro galvanized steel plate Chemical proof baking finish |
| Interior |
Stainless steel plate |
| Door |
Material of inside of chamber: Aluminum, without heat insulator, one touch handle mechanism |
| Door packing |
Vacuum O-ring (Fluororubber) |
| Heater |
5.25 kW |
| Temperature controller |
PID control with a micro computer (2-system heater control) |
| Temp. control/overheat preventive device sensor |
K-thermocouple |
| Sample temperature sensor |
K-thermocouple 6/chamber |
| Vacuum controller |
16 step pressure control, stable time setting function, vacuum control with timers (0 to 9999 sec.) |
| Vacuum setting |
On the touch panel |
| Vacuum valve |
Air driven, NW40 |
| Vacuum meter |
101 kPa to 10-2 Pa, vacuum indicator |
| Safety unit |
Over current ELB, independent overheat preventive device, stop valve, emergency relief valve, emergency stop valve, vacuum valve filter regulator |
| Standards |
Internal size |
W500 x D476 x H480 mm |
| External size |
W1100 x D850 x H1400 mm |
| Bottom surface effective size |
W480 x D420 x H480 mm |
| Exhaust port |
NW40 flange |
| Air supply port |
Rc1/2 |
| Power supply |
3 phase AC200V 50/60 Hz 16A (excluding vacuum pump power supply) |
| Compressed air |
0.5 Mpa or more, φ6 mm tube coupler is used. |