| Remarks: These Model Nos. are
not available in the U.S. market. |
| Model |
| 402 |
602 |
802 |
| Operating Temperature Range |
Room temp.
+5deg.C to 80 deg.C |
| Temp. Distribution Accuracy |
| ±1.0deg.C(at37deg.C) |
| Internal
Capacity |
| 97 L |
150 L |
318 L |
|
Incubators of natural convection by air-jacket,
equipped with various functions.
They are equipped with a newly developed controller,
and can be operated by an interactive key input, and are also equipped
with a temperature memory function.
- The air-jacket method makes it possible to uniformalize the
distribution of temperature.
- Setting can be made digitally using the dedicated operation
menu keys or the keys, setting, measuring and over-heat setting
temperature are displayed simultaneously.
- Samples can be observed through the inner door.
- Equipped with self-diagnosis function, a key-lock function
and an over current breaker.
- Possible to place the models IS402/602 in piles, one over
the other.
|
| Interior |

IC402/602 |

IC602 |
| |

IC802 |
| Control
Panel |
Shelf & Brackets |
|
 |
| Specifications |
| |
IC402 |
IC602 |
IC802 |
Performance
|
| Circulation method |
Natural convection
by air jacket |
| Operating temperature range |
Room temperature
+5 to 80deg.C |
| Temp. adjustment accuracy |
±0.5deg.C
(at 37deg.C) |
| Temp. distribution accuracy |
±1.0deg.C
(at 37deg.C) |
Components |
| Interior Material |
Stainless steel |
| Exterior material |
Electro-galvanized
steel sheet with melamine resin baking finish |
| Heat insulating material |
Glass wool |
| Heater |
Iron-chrome wire heater,
0.3KW |
Iron-chrome wire heater,
0.4KW |
Iron-chrome wire heater,
0.7KW |
Controller |
| Temp. controller |
PID control
by microprocessor |
| Temp. setting method |
Operation menu
key and digital setting by  key |
| Temp display method |
Measurement
temp. : Digital display by green LED
Setting temp. : Digital display by red LED |
| Timer |
1 min. to 99
Hrs. 59 min. and 100 Hrs. to 999 Hrs. 50min. |
| Operation function |
Fixed temperature
operation, quick auto-stop, Auto-stop, Auto-start |
| Additional functions |
Calibration
off-set, Key lock function, Uninterruptible power for memory |
| Heater circuit control |
SSR control |
| Sensor |
Pt sensor,
K-thermocouple |
Safe
Device |
| Safety countermeasures |
Self diagnosis functions (temp.
sensor abnormal, heater disconnection, SSR short, automatic
overheat prevention), Key lock function, Electric leakage breaker
with over current protection, Overheat prevention device |
Standard |
Internal dimensions (W×D×Hmm)
External dimensions(W×D×Hmm)
Internal capacity |
450×480×450
560×606×820
97 liters |
600×530×500
710×656×870
150 liters |
600×530×1,000
710×656×1,619
318 liters |
| Shelf plate with standard load |
Approx. 15
kg/piece |
| Shelf rest step number |
9 steps |
12 steps |
29 steps |
| Shelf rest pitch |
35 mm |
30mm |
| Power source 50/60Hz |
100V/220V, single phase,
3.2A/1.5A |
100V/220V, single phase,
4.2A/1.9A |
100V/220V, single phase,
7.5A/3.4A |
| weight |
Approx. 45 kg |
Approx. 65 kg |
Approx. 102 kg |
Accessories |
| Shelf plate |
Stainless steel, 2 pcs. |
Stainless steel, 2 pcs. |
Stainless steel, 4 pcs. |
| Shelf bracket |
4 pcs. |
4pcs. |
8 pcs. |
- The performance shown here is the value under the condition
of the ambient temperature of 23deg.C±5deg.C, humidity
of 65%RH±20%.
- The projection is not included for Internal dimensions and
External dimensions.
|
Optional
items |
| Remark |
Description |
Model No. |
Product Code |
| ○ |
Stand ON61 (for 402/602) |
IC Series |
211856 |
| |
Stacking Support OD40 (for 402) |
IC Series |
212822 |
| |
Stacking Support OD60 (for 602) |
IC Series |
212823 |
| |
Shelf (with support 2pcs.), ODK88 for 402 |
IC Series |
212246 |
| |
Shelf (with support 2pcs.), ODK22 for 602/802 |
IC Series |
212266 |
| |
Anti-vibration material with support EPM-08 |
IC Series |
851352 |
| |
Anti-vibration material without support EPM-05 |
IC Series |
851351 |
| ● |
25m dia Cable Port |
IC Series |
281121 |
| ● |
30m dia Cable Port |
IC Series |
281122 |
| ● |
Temperature Output Terminal 4-20mA, for ODK12 |
IC Series |
281123 |
| ● |
Time-up Output Terminal for ODK14 |
IC Series |
281124 |
|
Note:
1.
The item marked " ● " in the column of "Remark" shall be ordered together with the main unit.
2. The transportation cost will not be required if the order is together with main unit except the item marked "○".
In case you order the item marked "○" together with the main unit, please ask Yamato to obtain the correct transportation cost. |
|
| |
| Dimensions
(Unit:mm) |
 |
Contact us
Request reference material
Request for documents
Return to Home
Copyright(c) 1998 YAMATO SCIENTIFIC CO.,LTD. ALL rights