Neocool® Circulator   Space Saving Type
CF750/750SLabo Cube

Operating Temp. Range
-20 deg.C to room temp.
Cooling capacity
1,200W (at liquid temp.: 10 deg.C)

Slim, space-saving type that requires only minimum space and equipped with two circulation discharge systems and two solvent collection systems (CF750S)
This is an external sealing type cooling water circulator of space-saving type with a width of only 230 mm, requiring only minimum space when installing beside a laboratory table or a sink. Model CF750 is of space saving design with two sets of circulation discharge ports while CF750S is also of space saving design with two sets of circulation discharge ports as well as two sets of solvent collecting units as standard equipment, thus enabling cooling of two units using one circulator.

CF750/750S
CF750S / CF750
  • Employs non-Freon refrigerant R404A that conforms to the regulations on specific Freon.
  • Settings and display is made digitally on the operation part, which also has a timer function.
  • The circulation water flow adjusting cock, the discharge cock, and an overflow enables easy operation.
  • An over current electric leakage breaker, a refrigerator over load relay, a pump thermal protector, a delayed timer for refrigerator protection, and a H bypass for pump protection assure the optimal safety.
  • CF750S is equipped with a service socket for the vacuum pump.

Specifications
Product code 221519 221520
Model CF750 CF750S (WITH solvent collection unit)
System External sealed circulation
Performance
Operating temp. range -20 deg.C to room temp.
Temp. adjustment accuracy +/-2 deg.C
Cooling capacity Approx. 1200W (1030 Kcal/h) at liquid temp: 10 deg.C
Approx. 950W (820 Kcal/h) at liquid temp: 0 deg.C
Approx. 500W (430 Kcal/h) at liquid temp: -10 deg.C
Configuration
Temperature control Refrigerator ON-OFF control
Temperature sensor T-thermocouple (with SUS protective tube)
Temp and time setting Digital setting and LED display
Refrigerator/refrigerant Air-cooled 600W / R404A
Circulation pump With a magnet pump and an H bypass
Pump
performance
Max. flow
(50/60 Hz)
30 L/min
Max. lift
(50/60 Hz)
10.3/14.3m
Unit
circulation
performance
Max. flow
(50/60 Hz)
16.9/19.6 L/min
Max. lift
(50/60 Hz)
9.7/13.2m
Cooling coil Copper nickel plating
Operating environment temp. range 5 to 35 deg.C
Function
Safety functions Over current electrical leakage breaker, refrigerator over load relay, pump thermal protector, delay timer for protecting refrigerator, H bypass for protecting pump, drain, and over flow
External circulation nozzle O.D. 10.5 mm (SUS) I/O x two systems with stop valves
Solvent collection nozzle - O.D. 8 mm x two systems with stop valves; surface area: 850 cm2
Solvent collection mechanism - Systems with 500 mL collection bottles x 2
Standards
Power source AC100V, 15A (with service outlet 5A) 50/60 Hz
External dimensions 230 x 600 x 900 mm (W/D/H)
Weight Approx. 60 kg Approx. 65 kg
Accessory
Accessories Top cover, drain hose, operation manual, warranty card Solvent bottles x 2, top cover, drain hose, operation manual, warranty card



Installation example
Solvent collection unit for CF750S 2 systems

Control panel Cooling curve Flow/lift curve

Example system configuration (connection example between CF750S and RE)

Circulation heat insulation hose (optional) Vacuum hose (optional) Vacuum pump PG200 (optional)

Exhaust air flow 20/24 L/min (50/60 Hz)

Optional items
Description Model No. Product Code
Vacuum pump PG200 CF750/750S 255287
Circulation heat insulation hose dia. 9, 2m X 2 CF750/750S 221581
Vacuum hose dia. 6, 5m CF750/750S 255297

Dimensional drawing (mm)


Contact us

Index of Circulators

Return to home


Copyright(c) 1998 YAMATO SCIENTIFIC CO.,LTD. All rights reserved For more information, call 0120-405-525(Free dial) E-mail:info@yamato- net.co.jp